Summary
Using IMx Manager to configure a stand alone IMx provides limited configuration options. If a more complex configuration is required, this can be accomplished by first constructing the configuration in @ptitude Observer and then exporting it as a template that can be loaded to a stand alone IMx, from IMx Manager.
Details
Prerequisites
- An @ptitude Observer installation, version 10.5 or higher.
- An Enlight Centre account with IMx Manager access to a company.
Procedure
- In @ptitude Observer create an IMx and configure it with all the measurement points and communication options needed.
- From the IMx/MasCon devices dialog save this configuration using the Save Config button.
- Open Enlight Centre go to Hierarchy, then Company and drag the template to 'Additional files' to upload it.
- Open the IMx Manager app and select Settings then IMx Assets and verify that the templates are updated.
- When using IMx Manager to configure the target IMx as standalone, under Measurement Configuration choose 'template' and select the earlier created template from the list.
Contacting SKF Technical Support Group
For further assistance please open a support case using the Technical Support group's self-help portal at www.skf.com/cm/tsg. Once your support case is submitted, a technician will contact you to begin working on your issue. For urgent issues we are available at these times by phone:
- Monday through Friday, 5:00 a.m. to 4 p.m. Pacific Time -
Phone: +1 800 523 7514 within the US or +1 858 496 3627 outside the US. - Monday through Friday, 8:00 a.m. to 4:00 p.m. Central European Time -
Phone: +46 31 337 65 00. - Monday through Friday, 7:30 a.m. to 4:30 p.m. India Standard Time -
Phone: +60 16 699 9506.
Comments
0 comments
Please sign in to leave a comment.